Acta Orthopaedica, 78:4, 470-478

Uncemented metal-on-metal acetabular component

Uma T. Vassan, Sanjeev Sharma, Koganti P. Chowdary & Manjit S. Bhamra
Hip

Background Aseptic loosening is the major cause of implant failure. In cemented hip arthroplasty, failure of the acetabular side is mainly due to lysis caused by wear particles. By using an implant with low wear characteristics and by enhancing acetabular fixation using an uncemented implant, we aimed to reduce acetabular lysis and thereby loosening.

 

Patients and methods This was a retrospective cohort study of 119 hips (101 patients) that had the Uncemented Fitmore cup (Sulzer Orthopedics). In 66 patients, the femoral component was CF-30 (Sulzer) used with cement. In the remaining 35 patients, thrust plate prosthesis (TPP) (Sulzer) was used. Of the 101 patients, 94 (112 hips) were available for study. Mean follow-up of the 94 patients was 7 (5–13) years.

 

Results The mean preoperative Harris hip score was 38 and the mean postoperative Harris hip score was 89 at the last follow-up. Taking aseptic loosening as the endpoint, the survival rate of the Fitmore cup was 100% at 11 years.

 

Interpretation The uncemented Fitmore acetabular cup with second generation metal-on-metal articulation showed good results with regard to aseptic loosening in the medium term.


Link to article