The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 24, Issue: 5, Page: 751-758

Two-Year to Five-Year Follow-Up of Femoral Defects in Femoral Revision Treated With the Link MP Modular Stem

Rodriguez, Jose A; Fada, Robert; Murphy, Steven B; Rasquinha, Vijay J; Ranawat, Chitranjan S
Hip

Femoral deficiency has been shown to adversely affect the results of revision total hip arthroplasty. Tapered titanium modular stems allow distal fixation of the fluted, conical portion of the implant in the setting of proximal bone loss. One hundred two consecutive hips with proximal bone loss underwent revision femoral reconstruction between 1998 and 2002 at 3 centers using the Link MP modular stem. Forty-three hips had Mallory type 3C femoral deficiency. Ninety-seven hips were observed for an average of 45 months (range, 24-72; median, 36 months). Clinically, mean Harris hip score improved from 36 to 84 (range, 54-99). Radiographically, 93 hips were considered stable, with no circumferential lucencies at the distal fixation surface. Three hips migrated and required revision, along with one periprosthetic fracture. Five other hips had nonprogressive migration of 1 to 2 mm.


Link to article