International Orthopaedics April 2009, 33:403

Tibiofemoral contact areas and pressures in six high flexion knees

Shiramizu, K., Vizesi, F., Bruce, W. et al.
Knee

The tibiofemoral articulating interfaces of six high flexion knee designs were examined using a standard testing protocol developed by Harris et al. [J Biomech 32:951–958 (1999)] to investigate the polyethylene insert contact areas and pressures. A load of 3600 N was applied for 10 s at 0, 30, 60, 90, 110, 135 and 155° of flexion. Contact areas and pressures at the femoral–polyethylene insert interface were measured with a I-scan 4000 system. Up to 110°of flexion, the VANGUARD RP HI-FLEX showed the highest contact area and lowest pressure. At the deep flexion angles, contact area decreased and contact pressure increased significantly in all knees. The NexGen series showed a constant contact area throughout the various flexion angles. In general, all high flexion knees could result in almost point contact in an extremely high range of motion.