Acta Orthopaedica Scandinavica, 74:6, 677-682

The precision and influence of rotation for measurements of bone mineral density of the distal femur following total knee arthroplasty

Matthias Therbo, Michael Petersen, Henrik Schrøder, Palle Nielsen, Bo Zerahn & Bjarne Lund
Knee

We evaluated the feasibility of DEXA (Norland XR-26 mark II) for quantitative measurements of bone mineral density (BMD) in the lateral plane of the distal femur after total knee arthroplasty (TKA). BMD was measured in 5-6 regions of interest (ROI) in close relation to the femoral component. In an in vitro study using 3 different distal femur phantoms, we found that the precision was affected by rotation of the distal femur. When BMD measurements were repeated within a range of motion of 40°, 20°, and 0°, the coefficient of variation (CV) was approximately 15%, 10%, and 0.6%, respectively. We found that the use of bone cement for implant fixation had no effect on the level of BMD. Double measurements performed in 28 patients gave average CV values of 3.3%, 3.0%, and 2.6% for the uncemented Duracon, and Interax femoral components and the cemented AGC components, respectively. Our in vivo average CV measurements of BMD of the distal femur after TKA were on a level, suitable for repeated BMD measurements in prospective studies, wich evaluate adaptive bone remodeling of the distal femur after cemented and uncemented TKA.