Forty-six studies (12,520 knees) with an annual revision-rate ranging from 0% to 4.35% pa, mean 1.21% pa (95% confidence interval [CI], 0.97-1.47), were identified. In series with mean follow-up of 10-years, the revision-rate was 0.63% pa (95% CI, 0.46-0.83), equating to a 94% (95% CI, 92%-95%) 10-year survival. Aseptic loosening, lateral arthritis, bearing dislocation, and unexplained pain were the predominant failure mechanisms with revision for patellofemoral problems and polyethylene wear exceedingly rare. The lowest revision-rates were achieved with caseload >24 UKA/y (0.88% pa; 95% CI, 0.63-1.61) and usage >30% (0.69% pa; 95% CI, 0.50-0.90). Usage was more important than caseload; with high usage (≥20%), the revision-rate was low, whether the caseload was high (>12 UKA/y) or low (≤12 UKA/y; (0.94% pa; 95% CI, 0.69-1.23 and 0.85% pa; 95% CI, 0.65-1.08), respectively); with low usage (<20%), the revision-rate was high, whether the caseload was high or low (1.58% pa; 95% CI, 0.57-3.05 and 1.76% pa; 95% CI, 1.21-2.41, respectively).