The effects of femoral component design on the patello-femoral joint in a PS total knee arthroplasty. Arch Orthop Trauma Surg 134, 59–64 (2014).

The effects of femoral component design on the patello-femoral joint in a PS total knee arthroplasty

Indelli, P.F., Marcucci, M., Pipino, G. et al.
Knee

Introduction

Anterior knee pain following TKA performed utilizing the PFC Sigma system still represents a cause of failure. The purpose of this study was to evaluate whether or not a recent change in the femoral design (PFC Sigma PS) had a positive impact on the patello-femoral complication rate.

Materials and methods

A consecutive series of 100 TKA using the PFC Sigma PS system was followed prospectively for a minimum of 3 years. All patellae were replaced and a standard lateral release was never performed. Radiographic analysis following the Knee Society Score (KSS) included antero-posterior weight-bearing, lateral and bilateral axial radiographs. TKA rotational alignment was recorded at the final follow-up in 30 consecutive knees by performing a CT evaluation.

Results

Good to excellent clinical results according to the KSS were achieved in 94 % of the knees. Survival without need of reoperation for any reason was 98 % at 3 years minimum follow-up; two reoperations were done for removal of fibromatous intra-articular tissue (“Clunk syndrome”). There were no revisions for septic or aseptic loosening of the components. The mean ROM improved from 104° preoperatively to 115° (97°–132°) postoperatively: postoperative flexion was 120° or more in 58 % of the knees. Severe anterior knee pain was present in 9 % of patients. Radiographic evaluation showed 90 knees with a tibio-femoral anatomical axis between 8° and 2° of valgus (±3° from the intraoperative goal). CT evaluation of 30 consecutive knees showed that the femoral component positioning in relationship to the trans-epicondylar axis had only 2.80° of external rotation (±2.10°) with respect to a planned external rotation of 3°. This difference was statistically significant.

Conclusions

Although the PFC Sigma PS system provides good and predictable results for tricompartmental arthritis of the knee, anterior mechanism complications still represent a reason for dissatisfaction in a substantial group of patients.