The Journal of Arthroplasty, Volume 27, Issue 7, 1344 - 1348.e1

The 6-Year Migration Characteristics of a Hydroxyapatite-Coated Femoral Stem

Callary, Stuart A. et al.
Hip

A prospective cohort of 30 patients undergoing primary total hip arthroplasty for treatment of osteoarthritis was enrolled in a study to characterize the migration behavior of a clinically successful cementless stem. At 6 years, the mean subsidence of the stem was 0.63 mm (range, −0.33 to 3.68 mm); the mean rotation into retroversion was 1.41° (range, −1.33° to 7.48°). No stems had additional subsidence of more than 0.25 mm between 6 months and 6 years. The resultant mean subsidence between 2 and 6 years was 0.03 mm, which is below the limit measurable by radiostereometric analysis. The data demonstrate that subsidence of this cementless stem occurs within the first 6 months, after which there is persistent stabilization.