The Journal of Arthroplasty, Volume 29, Issue 7, 1443 - 1448

Tantalum Acetabular Augments in One-stage Exchange of Infected Total Hip Arthroplasty: A Case–control Study

Klatte, Till Orla et al.
Hip

During the one-stage exchange procedure for periprosthetic joint infection (PJI) after total hip arthroplasty (THA), acetabular defects challenge reconstructive options. Porous tantalum augments are an established tool for addressing acetabular destruction in aseptic cases, but their utility in septic exchange is unknown. This retrospective case–control study presents the initial results of tantalum augmentation during one-stage exchange for PJI. Primary endpoints were rates of re-infection and short-term complications associated with this technique. Study patients had no higher risk of re-infection with equivalent durability at early follow-up with a re-infection rate in both groups of 4%. In conclusion, tantalum augments are a viable option for addressing acetabular defects in one-stage exchange for septic THA. Further study is necessary to assess long-term durability when compared to traditional techniques for acetabular reconstruction.


Link to article