Reverse total shoulder arthroplasty using helical blade to optimize glenoid fixation and bone preservation: preliminary results in thirty five patients with minimum two year follow-up
Sebastien Zilber, Eleonora Camana, Peter Lapner, Emil Haritinian, Laurent Nove JosserandShoulder
Purpose
Glenoid loosening is a common cause of reverse total shoulder arthroplasty (RTSA) failure, and grafting of the glenoid is often required for revision due to bone loss due to the central peg in most glenoid baseplates. Helical blades have been used in the hip to optimize bone fixation in proximal femoral fracture. This study presents the initial results of specifically designed helical blade in the shoulder to optimize glenoid bone fixation and preservation as part of RTSA.
Methods
Thirty-five patients underwent RTSA with glenoid helical blade fixation. An uncemented glenoid baseplate was used with a central helical blade partially coated with hydroxyapatite and two or three screws. Outcome analysis was performed pre-operatively and at two years.
Results
All patients were satisfied with the results and significant improvement was observed in functional outcome scores between baseline and final follow-up. There was a single intra-operative undisplaced glenoid fracture which did not compromise the baseplate fixation. There was no radiographic evidence of loosening or radiolucencies around the helical blade.
Conclusions
The helical blade provides a satisfactory primary fixation. Because of its length (21 mm), care should be taken in cases of pre-existing bone loss or sclerotic bone to avoid glenoid fracture or anterior cortical perforation. Helical blade has the potential to facilitate glenoid implant revision by preserving the glenoid bone stock.
Link to article