Due to the well-documented problems surrounding metal-on-metal bearings, the use of hip resurfacing has declined. Since the potential benefits of hip resurfacing remain desirable, it may be beneficial to investigate the long-term outcome of hip resurfacings using metal-on-polyethylene in the 1980’s. We report the long-term survivorship and modes of failure of a cementless metal-on-polyethylene resurfacing (n = 178) with different porous ingrowth surfaces. While acetabular loosening was absent, a high incidence of femoral failures (femoral loosening = 18.1%, osteolytic neck fracture = 21%) occurred despite using the same ingrowth surface for both components. Ongoing developments using the lessons learned from these previous generation components and utilizing modern low wear materials, e.g., cross-linked polyethylene, may lead to improved implants for future hip resurfacings.
CoxaPro
> Clinical Library > Welcome to the joint replacement clinical library > Long-Term Outcome of a Metal-on-Polyethylene Cementless Hip Resurfacing
The Journal of Arthroplasty, Volume 29, Issue 4, 802 - 807
Hip
Long-Term Outcome of a Metal-on-Polyethylene Cementless Hip Resurfacing
Tan, Timothy L. et al.Hip