Clinical Orthopaedics and Related Research: December 2002 - Volume 405 - Issue - p 189-194

Late Dislocation of Rotating Platform in New Jersey Low-Contact Stress Knee Prosthesis

Huang, Chun-Hsiung MD*,†; Ma, Hon-Ming MD*; Liau, Jiann-Jong PhD†; Ho, Fang-Yuan MS**; Cheng, Cheng-Kung PhD†
Knee

Five patients with late rotational dislocation of the rotating platform bearing in the New Jersey Low-Contact Stress total knee arthroplasty are reported. The prostheses had functioned well for 8 to 12 years before failure. Preoperative radiographs showed asymmetric femorotibial joint spaces. Entrapment of the dislocated bearing in three patients and spontaneous reduction of the dislocated bearing in another two patients were seen at revision. Femorotibial ligamentous instability was found after reduction. The retrieved polyethylene bearings showed advanced wear and cold flow deformities and the thickness was reduced. The revision arthroplasty was accomplished by replacement with a thicker bearing element. Progressive femorotibial ligament laxity and reduction of the thickness of polyethylene with wearing break down the originally well-balanced soft tissue tension of the knee. The rotational degree of the rotating platform bearing is unrestricted, which may result in late dislocation. Polyethylene wear is unavoidable in knee prostheses using metal contact with polyethylene even with a mobile-bearing design. Efforts to reduce polyethylene wear are mandatory.