Intra- and post-operative accuracy assessments of two different patient-specific instrumentation systems for total knee replacement. Knee Surg Sports Traumatol Arthrosc 22, 621–629 (2014) doi:10.1007/s00167-013-2667-9

Intra- and post-operative accuracy assessments of two different patient-specific instrumentation systems for total knee replacement

Ensini, A., Timoncini, A., Cenni, F. et al.
Knee

Purpose

The aim of this study is to assess and compare the accuracy of two different patient-specific instrumentation (PSI) systems for total knee replacement, both intra-operatively for bone preparation and post-operatively for final component alignment.

 

Methods

Twenty-five patients were treated according to a computer tomography (CT)-based PSI system (group A) and 25 to a magnetic resonance imaging (MRI)/X-ray-based system (group B). Alignments on the three anatomical planes and resection thickness at the cutting blocks and at the resulting bone cuts were recorded intra-operatively by a standard surgical navigation system. Alignments of the prosthetic components and mechanical axis were also measured post-operatively on radiographs. These measurements at both the femur and tibia were compared with those of the corresponding pre-operative planning, considering discrepancies larger than 3° as outliers.

 

Results

In both groups, the mean absolute differences between pre-operatively planned alignments and corresponding intra- and post-operative measurements ranged from a minimum of 1.2° to a maximum of 2.9° in all three anatomical planes. In both groups and in both femur and tibia, the plane with the smallest percentage of outliers was the coronal, maximum 17 %. The comparison between two groups was statistically significant (p = 0.02) in the femoral sagittal plane, where group B showed smaller alignment discrepancies at the cutting blocks.

 

Conclusions

Both PSI systems showed good alignments in the coronal plane in all stages. For a few measurements, a better performance was observed in the MRI/X-ray-based system than in the CT-based system.

 

Level of evidence

I.