International Orthopaedics November 2016, Volume 40, Issue 11, pp 2295–2302

In-vivo kinematics of high-flex posterior-stabilized total knee prosthesis designed for Asian populations

Watanabe, T., Muneta, T., Koga, H. et al.
Knee

Purpose

The purpose of this study was to determine in-vivo kinematics of our developed posterior-stabilized (PS) total knee prosthesis for Asian populations in comparison with a popular high-flexion PS prosthesis.

Methods

We analyzed 62 osteoarthritic knees: 31 knees with the new PS prosthesis (group A) and 31 knees with a popular high-flexion PS prosthesis (group B). Radiographic knee images were taken during standing, lunge, and kneeling activities. The three-dimensional position and orientation of the implant components were determined using model-based shape matching techniques.

Results

Group A showed slightly greater implant flexion angles compared with knees with conventional prosthesis at maximum lunge (average: 119 vs. 110°, p = 0.001), and at maximum kneeling (121 vs. 114°, p = 0.004), although the range of motion was not significantly different. The femoral centre positions were more posterior in group A at standing, at 90° lunge, at maximum lunge (-9 and -7 mm, p = 0.004), at 90° kneeling, and at maximum kneeling (-9 vs. -7 mm, p = 0.016), and posterior translations of the femoral center were greater at 90° knee flexion postures. The femoral centre positions had a strong negative correlation with implant flexion angles at maximum lunge in group B (r = -0.893, p < 0.001), but not in group A (p = 0.242).

Conclusions

The new PS prosthesis designed for Asian knee morphology achieved flexion angles and range of motion at least comparable to that of conventional high-flexion PS prosthesis. The femoral roll-back pattern, however, is different from a conventional knee, reflecting the post/cam design.


Link to article