Accurate implant positioning and restoration of lower limb alignment are major requirements for successful long-term results in unicompartmental knee arthroplasty (UKA). Alignment accuracy was compared between navigated-UKA (nUKA) and conventional-UKA (cUKA) groups using a retrospective matched case–control study (n=129, 58 nUKA, 71 cUKA). Mechanical axis (MA), hip–knee–ankle angle (HKA°), coronal implant alignment, and tibial implant posterior slope were measured. No statistically significant difference was observed when comparing MA, HKA° or coronal implant alignment (p>0.05). Statistical significance was seen with tibial component posterior slope (p=0.04, nUKA 4.2°, cUKA 2.9°); and between intra-operative navigationally determined HKA° and post-operative whole-leg standing (WLS) film HKA°. Navigation does not significantly improve UKA alignment compared to conventional methods. Further studies are needed to justify the use of this technology in UKA.
CoxaPro
> Clinical Library > Welcome to the joint replacement clinical library > Implant and Limb Alignment Outcomes for Conventional and Navigated Unicompartmental Knee Arthroplasty
The Journal of Arthroplasty, Volume 28, Issue 3, 463 - 468
Knee
Implant and Limb Alignment Outcomes for Conventional and Navigated Unicompartmental Knee Arthroplasty
Valenzuela, Gustavo A. et al.Knee