The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 23, Issue: 4, Page: 573-580

How Does Electromagnetic Navigation Stack Up Against Infrared Navigation in Minimally Invasive Total Knee Arthroplasties?

Lionberger, David R; Weise, Jennifer; Ho, David M; Haddad, John L
Knee
Forty-six primary total knee arthroplasties were performed using either an electromagnetic (EM) or infrared (IR) navigation system. In this IRB-approved study, patients were evaluated clinically and for accuracy using spiral computed tomographic imaging and 36-in standing radiographs. Although EM navigation was subject to metal interference, it was not as drastic as line-of-sight interference with IR navigation. Mechanical alignment was ideal in 92.9% of EM and 90.0% of IR cases based on spiral computed tomographic imaging and 100% of EM and 95% of IR cases based on x-ray. Individual measurements of component varus/valgus and sagittal measurements showed EM to be equivalent to IR, with both systems producing subdegree accuracy in 95% of the readings.

Link to article