The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 36, Issue: 9, Page: 3333-3339

How Does Contact Length Impact Titanium Tapered Splined Stem Stability: A Biomechanical Matched Pair Cadaveric Study

Gkiatas, Ioannis; Malahias, Michael-Alexander; Xiang, William; Meyers, Kathleen N; Torres, Lisa A; Tarity, T David; Rodriguez, Jose A; Bostrom, Mathias P; Wright, Timothy M; Sculco, Peter K
Hip

Background

Titanium tapered stems (TTS) achieve fixation in the femoral diaphysis and are commonly used in revision total hip arthroplasty. The initial stability of a TTS is critical, but the minimum contact length needed and impact of implant-specific taper angles on axial stability are unknown. This biomechanical study was performed to better guide operative decision-making by addressing these clinical questions.

Methods

Two TTS with varying conical taper angles (2° spline taper vs 3.5° spline taper) were implanted in 9 right and left matched fresh human femora. The proximal femur was removed, and the remaining femoral diaphysis was prepared to allow for either a 2 cm (n = 6), 3 cm (n = 6), or 4 cm (n = 6) cortical contact length with each implanted stem. Stepwise axial load was then applied to a maximum of 2600N or until the femur fractured. Failure was defined as either subsidence >5 mm or femur fracture.

Results

All 6 femora with 2 cm of stem-cortical contact length failed axial testing, a significantly higher failure rate (P < .02) than the 4 out of 6 femora and all 6 femora that passed testing at 3 cm and 4 cm, respectively, which were not statistically different from each other (P = .12). Taper angle did not influence success rates, as each matched pair either succeeded or failed at the tested contact length.

Conclusion

4 cm of cortical contact length with a TTS demonstrates reliable initial axial stability, while 2 cm is insufficient regardless of taper angle. For 3 cm of cortical contact, successful initial fixation can be achieved in most cases with both taper angle designs.