Clinical Orthopaedics and Related Research: December 2004 - Volume 429 - Issue - p 63-67

Highly Cross-linked, Electron-Beam-Irradiated, Melted Polyethylene: Some Pros

Harris, William H MD, DSC
Hip

Extensive new evidence generated within the past the year provides strong support for the use of electron-beam highly cross-linked, subsequently melted ultra-high molecular weight polyethylene in total hip replacement arthroplasty. In terms of wear reduction, three studies involving three different demographic groups and two different measurement techniques have found that the femoral head penetration with this type of polyethylene after bedding in has finished taking place is less than 10 micra per year. This wear rate is similar to the wear rate of metal-on-metal and ceramic-on-ceramic articulations. Retrieval specimens up to 3 years after insertion confirmed the minimal wear by exhibiting persisting machine marks throughout the inside diameter of the liner. Extensive studies show no evidence of oxidation, confirming the absence of residual free radicals. No evidence of fatigue failure exists except three known cases out of 150,000, in which malposition of the acetabular component produced abnormally high contact stresses on unsupported polyethylene. The in vivo wear, oxidation resistance, and mechanical properties of this alternative bearing material are excellent, with in vivo durations now exceeding five years. The other major advantages over hard-on-hard bearings including familiarity, adaptability, forgiveness, and cost seem to be compelling.


Link to article