International Orthopaedics April 2018, Volume 42, Issue 4, pp 755–760

High bacterial contamination rate of electrocautery tips during total hip and knee arthroplasty

Hussein Abdelaziz, Akos Zahar, Christian Lausmann, Thorsten Gehrke, Helmut Fickenscher, Eduardo M. Suero, Matthias Gebauer, Mustafa Citak
Hip Knee

Purpose

The aim of the study was to quantify the bacterial contamination rate of electrocautery tips during primary total joint replacement (TJR), as well as during aseptic and septic revision TJR.

Methods

A total of 150 electrocautery tips were collected between April and July 2017. TJR surgeries were divided into three groups: (1) primary, (2) aseptic and (3) septic revisions. In each group, a total of 50 electrocautery tips were collected. A monopolar electrocautery with a reusable stainless-steel blade tip was used in all cases. The rate of bacterial contamination was determined for all groups. Correlation of exposure time and type of surgery was analyzed.

Results

The overall bacterial contamination rate was 14.7% (95% CI 9.4 to 21.4%). The highest contamination rate occurred in the septic revision group (30.0%; 95% CI 17.9 to 44.6%), followed by the primary cases group (10.0%; 95% CI 3.3 to 21.8%) and the aseptic revision group (4.0%; 95% CI 0.5 to 13.7%). Exposure time did not affect the bacterial contamination rate. In 12 out of 15 (80%) contaminations identified in the septic group, we found the same causative microorganism of the prosthetic joint infection on the electrocautery tip.

Conclusions

The bacterial contamination of the electrocautery tips is relatively high, especially during septic hip revision arthroplasty. Electrocautery tips should be changed after debridement of infected tissue.