CoxaPro
> Clinical Library > Welcome to the joint replacement clinical library > Femoral Shaft Fracture Resulting From Femoral Tracker Placement in Navigated TKA
Clinical Orthopaedics and Related Research: June 2008 - Volume 466 - Issue 6 - p 1499-1502 doi: 10.1007/s11999-008-0150-6
Knee
Link to article
Femoral Shaft Fracture Resulting From Femoral Tracker Placement in Navigated TKA
Bonutti, Peter1; Dethmers, Daniel1; Stiehl, James, B.2,3,aKnee
Computer-assisted navigation is a surgical tool that may decrease malalignment outliers in TKA. With any new surgical technique, there is the possibility of unexpected complications that raise caution. We report two patients with displaced femoral fractures at optical tracker pin placement sites created for routine performance of navigated TKA. Our experience suggests single bicortical 5-mm pins placed in the femoral shaft have the added risk of creating a stress riser leading to the potential for fracture. Females may have a higher risk for this complication. We believe bicortical pin fixation in the femur or tibia no longer is indicated.
Link to article