The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 22, Issue: 2, Page: 293-296

Extensive Intrapelvic Granuloma Formation Caused by Ceramic Fragments After Revision Total Hip Arthroplasty

Kohn, Dieter; Pape, Dietrich
Hip

Ceramic debris embedded in a polyethylene liner was identified as the cause for rapid metal wear after revision total hip arthroplasty when a broken ceramic head was replaced by a metal head. Removal of an intrapelvic foreign body granuloma and rerevision hip arthroplasty had to be performed within 2 years after the first revision.


Link to article