HIP International. 2021;31(6):735-742.

Disappointing long-term outcome of THA with carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) as acetabular insert liner: a prospective study with a mean follow-up of 14.3 years

Heijnens LJ, Schotanus MG, Verburg AD, van Haaren EH.
Hip

Insert liner wear of the acetabular component is one of the predictive values for survival of total hip arthroplasties (THAs). This prospective single-centre study was designed to evaluate the follow-up of carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) insert liner used as bearing in cementless THAs.

29 healthy patients with an indication for cementless THA were selected for a CFR-PEEK insert liner and followed over time. All patients received a cementless THA with a CFR-PEEK insert liner used as bearing. At different follow-up moments patients were routinely examined and were analysed using the Oxford Hip Score (OHS), the modified Merle d’Aubigne-Postel (MAP) score, and radiologically. At the follow up moments the plain radiographics where assessed for loosening, cyst formations and wear of the CFR-PEEK liners.

At a mean of 14.3 years follow-up 4 revisions of the acetabular component were performed, resulting in a survival rate of 86.5% (CI 95%, 72.4–96.6). A statistically significant difference in OHS and MAP scores between pre- and postoperative follow-up moments was observed. The acetabular components of the remaining patients showed no radiological abnormalities at 14.3 years follow-up. The overall CFR-PEEK wear was low, with a mean of 0.81 (0.2–1.4) mm wear at 14.3 years follow-up.

In this series we found an aseptic loosening with unclear reasons in 4 well-positioned acetabular components, hence we do not recommend routine use of CFR-PEEK insert liners as bearing in cementless THAs. All the remaining THAs and acetabular components were in situ without abnormalities at 14.3 years follow-up.