Clinical Orthopaedics and Related Research: August 2005 - Volume 437 - Issue - p 132-137

Cup Alignment Error Model for Total Hip Arthroplasty

Wolf, Alon PHD; DiGioia, Anthony M III MD; Mor, Andrew B PHD; Jaramaz, Branislav PHD
Hip

Almost all computer-assisted orthopaedic surgery systems that rely on the anterior pelvic plane definition, such as in computed tomography and magnetic resonance image-based, fluoroscopy-based, and nonimage total hip replacement approaches, are derived from identifying two pairs of pelvic bony landmarks: anterior superior iliac spines and the pubic tubercles. Although these systems strive to achieve cup alignment accuracy of approximately 1°, even a minor failure to correctly identify these anatomic landmarks can lead to higher inaccuracies in the final cup alignment. This study shows how to examine the effects of these inaccuracies on the final acetabular cup implant orientation during total hip replacement by generating a kinematic model, which then is simulated. Simulation results indicate that, for example, a total error of 4 mm in measuring the anterior superior iliac spine and the pubic tubercles would result in a final cup orientation of 47° and 27° in abduction and version respectively, resulting in a 2° abduction error and 7° error in version when targeting 45° abduction and 20° version results. These calculations can be repeated for any error values.


Link to article