Clinical Orthopaedics and Related Research: December 2012 - Volume 470 - Issue 12 - p 3537–3541 doi: 10.1007/s11999-012-2574-2 Clinical Research

Cross-sectional Anatomy of the Ilium: Implications for Acetabular Component Placement in Total Hip Arthroplasty

Antoniades, John, MD1, a; Pellegrini, Vincent, D., Jr, MD1
Hip

Background High hip center reconstructions, used in revision and complex primary THAs, rely on pelvic bone stock at least 35 mm above the anatomic teardrop. However, the technique does not restore normal hip biomechanics and controversy exists regarding acetabular implant survival. Previous reports document a wide range of implant positioning above the teardrop. There is no anatomic guidance in the literature regarding the amount of bone stock available for initial implant stability in this area of the ilium.

 

Questions/purposes We therefore determined the thickness of the human ilium and related it to acetabulum cup coverage in high hip center reconstructions.

 

Methods We sectioned 16 cadaveric hips from the anterior superior iliac spine to the anatomic teardrop in 5-mm increments, then measured the thickness of the ilium for each cross section.

 

Results The maximum thickness of 42 ± 9 mm occurred at the dome of the acetabulum 35 ± 3 mm above the teardrop. At a distance of 1 cm above the dome, the ilium was reduced by 24%, to 32 ± 6 mm. At 2 cm above the dome, the ilium thickness was 22 ± 4 mm, a 48% reduction from its maximum.

 

Conclusion There are substantial anatomic limitations to high hip reconstructions 2 cm above the acetabular dome.