The Journal of Arthroplasty, Volume 35, Issue 6, S289 - S293

Clinical Results of Revision Hip Arthroplasty for Neck-Taper Corrosion and Adverse Local Tissue Reactions Around a Modular Neck Stem

Inoue, Daisuke et al.
Hip

Background

Adverse local tissue reactions (ALTRs) around a modular neck stem at our institution lead to a 13.5% rate of revision. The purpose of this study was to report the clinical results of revision total hip arthroplasty (THA) in this patient population.

Methods

We identified 80 hips in 77 patients who underwent revision THA due to neck-stem corrosion. Intraoperative and postoperative complications, clinical outcomes, re-revision rates, and the postoperative ion levels were recorded.

Results

The mean follow-up period after revision was 45.3 ± 16.3 months (range 24-81 months). There were no intraoperative mechanical complications during the revision surgery. Eight hips (10%) had postoperative complications related to the revision implant, of which 6/8 came to re-revision: postoperative implant dislocation in 4 hips—2 treated nonsurgically, aseptic loosening of femoral component in 2 hips, and periprosthetic femoral fracture in 2 hips. No patients had recurrence of ALTR nor the recurrence of groin pain within the follow-up period. With 2 exceptions, ion levels normalized within one year of the revision surgery. The survival rate was 94% (95% confidence interval 84-98) at 32 months and 82% (95% confidence interval 56-93) at 63 months after revision THA (using re-revision THA due to any reason as the endpoint).

Conclusion

Intraoperative complication, postoperative complication, and re-revision rates are equivalent to femoral revision surgeries for other causes (infection, fracture, loosening). We recommend selecting revision-type stems to minimize the risk of femoral loosening or periprosthetic fracture, and larger femoral heads or dual mobility bearings to minimize the risk of dislocation.