At the most recent follow-up, 5 patients had died, 3 patients have been revised because of ongoing instability, and 3 patients have had revision surgery due to infection. Overall functional outcome (mean Western Ontario and McMaster Universities Osteoarthritis Indexfunction 76, University of California, Los Angeles 5.6, mean Oxford 74.7, Short Form-12 physical 41.6/mental 53.3) and overall pain relief (mean Western Ontario and McMaster Universities Osteoarthritis Index pain score 78.3) scores were good. The mean satisfaction score was 78 of 100. The median serum trace metal chromium and cobalt levels at the most recent follow-up were 0.4 µg/L (range 0.1-6.1 µg/L) and 0.42 µg/L (range 0.21-9.42 µg/L), respectively. The survival with revision as the end point was 90%.