The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 24, Issue: 5, Page: 815-818

Acute Liner Disassociation of a Pinnacle Acetabular Component

J. Wesley Mesko
Hip

A primary total hip arthroplasty using a cementless Pinnacle cup and neutral Marathon polyethylene liner mated via a 32-mm cobalt chrome head to a Summit pore coated stem (Depuy, Warsaw, IN). At 23 months postoperative an acute disassociation occurred. This was resolved with a liner exchange to a 36-mm metal liner with a 36-mm metal head. The retrieved polyethylene liner was noted to have 3 of the 6 peripheral locking tabs sheared off.