J Orthop Surg Res 13, 275 (2018).

Accurate determination of post-operative 3D component positioning in total knee arthroplasty: the AURORA protocol

Wakelin, E.A., Tran, L., Twiggs, J.G. et al.
Knee

Background

Successful component alignment is a major metric of success in total knee arthroplasty. Component translational placement, however, is less well reported despite being shown to affect patient outcomes. CT scans and planar X-rays are routinely used to report alignment but do not report measurements as precisely or accurately as modern navigation systems can deliver, or with reference to the pre-operative anatomy.

Methods

A method is presented here that utilises a CT scan obtained for pre-operative planning and a post-operative CT scan for analysis to recreate a computation model of the knee with patient-specific axes. This model is then used to determine the post-operative component position in 3D space.

Results

Two subjects were investigated for reproducibility producing 12 sets of results. The maximum error using this technique was 0.9° ± 0.6° in rotation and 0.5 mm ± 0.3 mm in translation. Eleven subjects were investigated for reliability producing 22 sets of results. The intra-class correlation coefficient for each of the three axes of rotation and three primary resection planes was > 0.93 indicating excellent reliability.

Conclusions

Routine use of this analysis will allow surgeons and engineers to better understand the effect of component alignment as well as the placement on outcome.


Link to article