International Orthopaedics June 2015, Volume 39, Issue 6, pp 1057–1064

A novel non-invasive mechanical technique of cup and stem placement and leg length adjustment in total hip arthroplasty for dysplastic hips

Murayama, T., Ohnishi, H., Mori, T. et al.
Hip

Purpose

This study assessed the accuracy of cup and stem positioning and limb length adjustment for developmental dysplasia of the hip (DDH) using our new mechanical technique compared with imageless navigation or a computed tomography (CT)-based navigation system.

Methods

One hundred thirteen primary total hip arthroplasties (THAs) for DDH were evaluated. At pre-operative positioning, patients were placed in a precise lateral decubitus position by tilting the surgical table using simple ready-made devices (two shot pipe, metal chain, level gauge and goniometer). During surgery, cups were intentionally placed at 45° inclination and 15° anteversion on radiograph by using a level gauge and goniometer.

Results

Cup inclination was 44.2° ± 3.4° (range, 32.0–51.2°), cup anteversion was 19.6° ± 6.1° (range, 3.0–33.1°), stem alignment was 0.04° ± 0.8° valgus (range, 2.1° varus to 1.9° valgus), and leg length discrepancy was −0.37 ± 3.7 mm (range, −12.8 to 8.8 mm) in postoperative radiographs. Outliers (outside ±10° from intentional position) occurred in 15 cases (13.3 %) in inclination or anteversion. Postoperative dislocation did not occur in any cases.

Conclusions

Cup and stem positioning in THAs with our new mechanical technique yielded satisfactory results compared with previously reported imageless navigation or CT-based navigation. Our results were superior with regard to being non-invasive and low cost and involving minimum radiation exposure.


Link to article