The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 16, Issue: 5, Page: 658-665

A new pin-on-disk wear testing method for simulating wear of polyethylene on cobalt-chrome alloy in total hip arthroplasty

Charles R. Bragdon; Daniel O. O'Connor; Jay D. Lowenstein; Murali Jasty; Stephen A. Biggs; William H. Harris
Hip
Abstract: Hip simulator studies show that the wear of ultra-high molecular weight polyethylene against a cobalt alloy head depends on the wear path, especially the combination of a predominantly linear wear direction on which is superimposed motions in different directions. We postulated that multidirectional motion was necessary to generate realistic wear rates in pin-on-disk testing. To assess this hypothesis, a new pin-on-disk tester was developed, capable of unidirectional and bidirectional motion. Unidirectional motion produced no detectable wear. The rectangular motion produced wear rates, surface morphologies, and wear particles consistent with human acetabular specimens. The results for 1 Hz and 2 Hz were similar.

Link to article