Clinical Orthopaedics and Related Research: February 2011 - Volume 469 - Issue 2 - p 395–404 doi: 10.1007/s11999-010-1555-6 Symposium: Papers Presented at the Hip Society Meetings 2010

Wear versus Thickness and Other Features of 5-Mrad Crosslinked UHMWPE Acetabular Liners

Shen, Fu-Wen, PhD1; Lu, Zhen, PhD1; McKellop, Harry, A., PhD1, a
Hip

Background The low wear rates of crosslinked polyethylenes provide the potential to use larger diameters to resist dislocation. However, this requires the use of thinner liners in the acetabular component, with concern that higher contact stresses will increase wear, offsetting the benefits of the crosslinking.

 

Questions/purposes We asked the following questions: Is the wear of conventional and crosslinked polyethylene liners affected by ball diameter, rigidity of backing, and liner thickness? Are the stresses in the liner affected by thickness?

 

Methods Wear rates were measured in a hip simulator and stresses were calculated using finite element modeling.

 

Results Without crosslinking, the wear rate was 4% to 10% greater with a 36-mm diameter than a 28-mm diameter. With crosslinking, wear was 9% lower with a 36-mm diameter without metal backing and 4% greater with metal backing. Reducing the thickness from 6 mm to 3 mm increased the contact stress by 46%, but the wear rate decreased by 19%.

 

Conclusions The reduction in wear with 5 Mrad of crosslinking was not offset by increasing the diameter from 28 mm to 36 mm or by using a liner as thin as 3 mm.

 

Clinical Relevance The results indicate, for a properly positioned 5-Mrad crosslinked acetabular component and within the range of dimensions evaluated, neither wear nor stresses in the polyethylene are limiting factors in the use of larger-diameter, thinner cups to resist dislocation.