The Journal Of Bone And Joint Surgery - Volume 99 - Issue 17 - p. 1447-1452

Wear of Vitamin E-Infused Highly Cross-Linked Polyethylene at Five Years

Shareghi Bita, BSc; Johanson Per-erik, MD; Kärrholm Johan, MD, PhD
Hip
Background: In an earlier study with a 2-year follow-up of uncemented cups, we had reported low femoral-head penetration of vitamin E-infused highly cross-linked polyethylene liners (E1) compared with highly cross-linked liners without vitamin E (ArComXL). We studied the penetration rate of E1 compared with that of ArComXL, with a focus on changes occurring between 2 and 5 years after total hip arthroplasty.
Methods: In this randomized controlled study, we performed radiostereometric analysis of the penetration rate up to 5 years in 63 hips.
Results: During the total period of observation, the median proximal penetration for E1, 0.13 mm (mean, 0.11 mm [95% confidence interval (CI), 0.08 to 0.14 mm]), was lower than that for ArComXL, 0.20 mm (mean, 0.22 mm [95% CI, 0.17 to 0.26 mm]). The median proximal penetration rate between 2 and 5 years was 0.02 mm/yr (mean, 0.01 mm/yr [95% CI, 0.01 to 0.02 mm/yr]) for E1 and 0.04 mm/yr (mean, 0.04 mm/yr [95% CI, 0.03 to 0.05 mm/yr) for ArComXL. The corresponding median total (i.e., 3-dimensional resultant) penetration rates were 0.04 mm/yr (mean, 0.04 mm/yr [95% CI, 0.03 to 0.05 mm/yr]) for E1 and 0.07 mm/yr (mean, 0.08 mm/yr [95% CI, 0.06 to 0.10 mm/yr]) for ArComXL.
Conclusions: From years 2 to 5, we observed increased penetration in both groups. The penetration rate was higher for ArComXL, resulting in more proximal and total penetration at 5 years than for E1.
Level of Evidence: Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.