Bone Joint J. 2020 Oct; 102-B(10): 1303–1310.

Vitamin E-doped total hip arthroplasty liners show similar head penetration to highly cross-linked polyethylene at five years: a multi-arm randomized controlled trial

Kristian Kjærgaard, MD, Research Assistant,1,2 Ming Ding, MD, PhD, DMSci, Professor,1,2 Carsten Jensen, MSci, PhD, Associate Professor,3,4 Charles Bragdon, MSci, PhD, Associate Professor,5 Henrik Malchau, MD, PhD, Professor,5 Christina M. Andreasen, MSci, PhD, Assistant Professor,1,2 Ole Ovesen, MD, Consultant Orthopaedic Surgeon,1 Christian Hofbauer, MD, Consultant Orthopaedic Surgeon,6 and Søren Overgaard, MD, DMSci, Professor, Consultant Orthopaedic Surgeon1,2
Hip

Aims

The most frequent indication for revision surgery in total hip arthroplasty (THA) is aseptic loosening. Aseptic loosening is associated with polyethylene liner wear, and wear may be reduced by using vitamin E-doped liners. The primary objective of this study was to compare proximal femoral head penetration into the liner between a) two cross-linked polyethylene (XLPE) liners (vitamin E-doped (vE-PE)) versus standard XLPE liners, and b) two modular femoral head diameters (32 mm and 36 mm).

Methods

Patients scheduled for a THA were randomized to receive a vE-PE or XLPE liner with a 32 mm or 36 mm metal head (four intervention groups in a 2 × 2 factorial design). Head penetration and acetabular component migration were measured using radiostereometric analysis at baseline, three, 12, 24, and 60 months postoperatively. The Harris Hip Score, University of California, Los Angeles (UCLA) Activity Score, EuroQol five-dimension questionnaire (EQ-5D), and 36-Item Short-Form Health Survey questionnaire (SF-36) were assessed at baseline, three, 12, 36, and 60 months.

Results

Of 220 screened patients, 127 were included in this study. In all, 116 received the allocated intervention, and 94 had their results analyzed at five years. Head penetration was similar between liner materials and head sizes at five years, vE-PE versus XLPE was -0.084 mm (95% confidence interval (CI) -0.173 to 0.005; p = 0.064), and 32 mm versus 36 mm was -0.020 mm (95% CI -0.110 to 0.071; p = 0.671), respectively. No differences were found in acetabular component migration or in the patient-reported outcome measures.

Conclusion

No significant difference in head penetration was found at five years between vE-PE and XLPE liners, nor between 32 mm and 36 mm heads.


Link to article