Visual intraoperative estimation of range of motion is misleading in minimally invasive total hip arthroplasty. Arch Orthop Trauma Surg 136, 1015–1020 (2016).

Visual intraoperative estimation of range of motion is misleading in minimally invasive total hip arthroplasty

Woerner, M., Weber, M., Sendtner, E. et al.
Hip

Introduction

Generally range of motion (ROM) in total hip arthroplasty (THA) is intraoperatively assessed by eye. Can we assume that visual estimation of ROM is reliable?

Methods

60 patients underwent cementless THA in a subgroup analysis of a clinical prospective trial using a minimally invasive anterolateral approach in lateral decubitus position. Four experienced surgeons intraoperatively estimated ROM visually by assessment of the femur relative to the alignment of the patient’s pelvis. These estimations were compared with computer navigation measurements.

Results

We found a mean difference between navigation measurements and intraoperative estimations by eye of −5.6° (±10.9°; −17° to 30°) for flexion, respectively, −0.4° (±10.7°; −24° to 30°) for extension, 8.7° (±9.0°; −10° to 34°) for abduction, 5.9° (±18.3°; −58° to 68°) for external rotation and −5.8° (±12.1°; −38° to 22°) for internal rotation. Multivariate analysis showed no association between the visual accuracy of estimation of ROM and patient characteristics, such as BMI, sex, grade of osteoarthritis and treatment side except for a significant correlation of visual accuracy of estimation of extension and the level of professional experience. Otherwise, the level of professional experience had no impact on the accuracy of estimation of ROM by eye.

Conclusions

Even the experienced surgeon’s intraoperative estimation of ROM by eye is not reliable and differs up to 30° compared to objective measurements in minimally invasive THA. For accurate intraoperative assessment of ROM, the use of technical devices is recommended.