The Journal of Arthroplasty, Volume 35, Issue 10, 3010 - 3030

Videofluoroscopic Evaluation of the Influence of a Gradually Reducing Femoral Radius on Joint Kinematics During Daily Activities in Total Knee Arthroplasty

Renate List, Ruediger von Eisenhart-Rothe et al.
Knee

Background

Paradoxical anterior translation in midflexion is reduced in total knee arthroplasties (TKAs) with a gradually reducing femoral radius, when compared to a 2-radii design. This reduction has been shown in finite element model simulations, in vitro tests, intraoperatively, and recently also in vivo during a lunge and unloaded flexion-extension. However, TKA kinematics are task dependent and this reduction has not been tested for gait activities.

Methods

Thirty good outcome subjects (≥1 year postoperatively) with a unilateral cruciate-retaining TKA with a gradually reducing (n = 15) or dual (n = 15) femoral radius design were assessed during 5 complete cycles of level walking, stair descent (0.18-m steps), deep knee bend, and sitting down onto and standing up from a chair, using a moving fluoroscope (25 Hz, 1 ms shutter time). Kinematic data were extracted by 2D/3D image registration.

Results

Tibiofemoral ranges of motion for flexion-extension, abduction-adduction, internal-external rotation, and anteroposterior (AP) translation were similar for both groups, whereas the pattern of AP translation-flexion-coupling differed. The subjects with the dual-radii design showed a sudden change in direction of AP translation around 30° of flexion, which was not present in the subjects with the gradually reducing femoral radius design.

Conclusion

Through the unique ability of moving fluoroscopy, the present study confirmed that the gradually reducing femoral radii eliminated the paradoxical sudden anterior translation at 30° present in the dual-radii design in vivo during daily activities, including gait and stair descent.

Link to article