The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 22, Issue: 6, Page: 849-858

Unconstrained Tripolar Implants for Primary Total Hip Arthroplasty in Patients at Risk for Dislocation

Guyen, Olivier; Pibarot, Vincent; Vaz, Gualter; Chevillotte, Christophe; Carret, Jean-Paul; Bejui-Hugues, Jacques
Hip

We performed a retrospective study on 167 primary total hip arthroplasty (THA) procedures in 163 patients at high risk for instability to assess the reliability of unconstrained tripolar implants (press-fit outer metal shell articulating a bipolar polyethylene component) in preventing dislocations. Eighty-four percent of the patients had at least 2 risk factors for dislocation. The mean follow-up length was 40.2 months. No dislocation was observed. Harris hip scores improved significantly. Six hips were revised, and no aseptic loosening of the cup was observed. The tripolar implant was extremely successful in achieving stability. However, because of the current lack of data documenting polyethylene wear at additional bearing, the routine use of tripolar implants in primary THA is discouraged and should be considered at the present time only for selected patients at high risk for dislocation and with limited activities.