Clinical Orthopaedics and Related Research: May 2001 - Volume 386 - Issue - p 235-242

Tibiofemoral Conformity and Kinematics of Rotating-Bearing Knee Prostheses

D’Lima, Darryl D. MD; Trice, Michael MD; Urquhart, Andrew G. MD; Colwell, Clifford W. Jr. MD
Knee

Increasing tibiofemoral articular conformity theoretically increases articular contact area and reduces contact stresses in total knee arthroplasty. Fixed-bearing knee designs possess relatively low tibiofemoral conformity, in part to allow tibiofemoral rotation without generating excessive stresses at the articulation or the implant-bone interface. This study analyzed knee kinematics of mobile-bearing designs in a closed chain dynamic knee extension model in posterior cruciate-retaining design with high-and low tibiofemoral conformity and posterior cruciate-substituting designs with and without rotational constraint. Overall, for all conditions, the mobile-bearing insert rotated with the femur in the presence of tibiofemoral axial rotation. In addition, the correlation of bearing rotation with femoral rotation was stronger for the high-conformity and rotationally-constrained designs than for the low-conformity designs and strongest for the posterior cruciate-retaining high-conformity condition. Changes in conformity or rotational constraint did not appear to affect femoral roll back, tibiofemoral axial rotation, or varus-valgus angulation. The results suggest that mobile-bearing inserts rotate with the femur and increasing conformity or rotational constraint in mobile-bearing design knee prostheses does not affect knee kinematics adversely, at least under closed chain knee extension conditions in vitro.