The Journal of Arthroplasty, Volume 30, Issue 7, 1260 - 1264

Three Year RSA Evaluation of Vitamin E Diffused Highly Cross-linked Polyethylene Liners and Cup Stability

Sillesen, Nanna H. et al.
Hip

Vitamin E diffusion into highly cross-linked polyethylene (E-XLPE) is a method for enhancing oxidative stability of acetabular liners. The purpose of this study was to evaluate in vivo penetration of E-XLPE using radiostereometric analysis (RSA). Eighty-four hips were recruited into a prospective 10-year RSA. This is the first evaluation of the multicenter cohort after 3-years. All patients received E-XLPE liners (E1, Biomet) and porous-titanium coated cups (Regenerex, Biomet). There was no difference (P = 0.450) in median femoral head penetration into the E-XLPE liners at 3-years comparing cobalt-chrome heads (−0.028 mm; inter-quartile range (IQR) — 0.065 to 0.047) with ceramic heads (−0.043 mm, IQR — 0.143 to 0.042). The 3-year follow-up indicates minimal E-XLPE liner penetration regardless of head material and minimal early cup movement.


Link to article