The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 36, Issue: 10, Page: 3507-3512

Survivorship of Ceramic-on-Ceramic Total Hip Arthroplasty With Metal-Backed Acetabular Liners at 10 Years

Wagner, Moritz; Schönthaler, Hannes; Endstrasser, Franz; Neururer, Sabrina; Leitner, Hermann; Brunner, Alexander
Hip

Background

Ceramic-on-ceramic bearings are becoming increasingly popular in primary total hip arthroplasty (THA). To enhance ceramic-on-ceramic liner exchange in case of revision surgery, metal-backed liner systems have been proposed. Little is known about the clinical performance of these implants. The purpose of this study is to evaluate a metal-backed liner implant system for primary THA.

Methods

A total of 422 patients (with 468 consecutive THAs) were followed over a mean period of 10 years. All arthroplasties were performed with a cementless stem, a press-fit cup, and a metal-backed liner system. Surgical and clinical data, complications, and revisions were analyzed. The Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) before surgery and at 1 and 10 years after surgery were compared.

Results

The overall 10-year implant survival rate was 93.8%. The survival rate was 97.0% for heads and liners, 97.5% for stem, and 99.3% for acetabular cup. The most common reason for revision was ceramic breakage (2.4%) of the third-generation (BIOLOX forte) acetabular liner. Mean WOMAC score improved significantly from 50.1 before surgery to 13.2 at 1 year after surgery. There was no difference in WOMAC scores between surgical approach and type of bearing at 1 and 10 years after surgery.

Conclusion

THA using cementless stem, press-fit cup, and metal-backed liner system provides satisfactory long-term outcomes, with revision rate comparable to that with other systems available in the market. The metal-backed liner system has low risk of mal-seating. Third-generation ceramic liners should be avoided as they seem to be more prone to breakage.