The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 20, Issue: 3, Page: 386-391

Stress Risers Between Two Ipsilateral Intramedullary Stems

Iesaka, Kazuho; Kummer, Frederick J; Di Cesare, Paul E
Hip
Periprosthetic fracture after total joint replacement predominantly occurs at the stem tip. In this study, the effects of gap size, stem stability and cortical thickness between two press-fit, ipsilateral intramedullary stems on the tensile stresses created in the femur were investigated using finite-element models. The findings were confirmed with strain-gauge tests using a composite Sawbone femur. Gap size did not affect the level of stress on the femur. Cortical thickness had an important effect on stress distribution: peak stresses increased as bone cortical thickness decreased. Irrespective of gap size, the tips of loose stems acted as stress risers particularly with thinner cortices; the tips of well-fixed stems, however, did not.

Link to article