International Orthopaedics January 2010, Volume 34, Issue 1, pp 19–26

Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty

Gallo, J., Havranek, V. & Zapletalova, J.
Hip

We analysed data from 155 revisions of identical cementless hip prostheses to determine the influence of patient-, implant- and surgery-related factors on the polyethylene wear rate and size of periprosthetic osteolysis (OL). This was calculated by logistic regression analysis. Factors associated with an increased/decreased wear rate included position of the cup relative to Kohler’s line, increase in abduction angle of the cup, traumatic and inflammatory arthritis as a primary diagnosis, and patient height. Severe acetabular bone defects were predicted by an increased wear rate (odds ratio, OR = 5.782 for wear rate above 200 mm3/y), and increased height of the patient (OR = 0.905 per each centimetre). Predictors of severe bone defects in the femur were the increased wear rate (OR = 3.479 for wear rate above 200 mm3/y) and placement of the cup outside of the true acetabulum (OR = 3.292). Variables related to surgical technique were the most predictive of polyethylene wear rate.