Our study demonstrated comparable 3-D linear penetration rates, which were 0.06 ± 0.07 mm/y for group S and 0.03 ± 0.02 mm/y for group L at 10 years after surgery and penetration rates seemed to be almost constant with no significant difference after 3 years. Minimum liner thickness (5.3 mm for 48-mm socket in combination with 32-mm femoral head) and the second thinnest XLPE (6.3 mm in case of socket from 50 mm to 54 mm combined with 32-mm femoral head) was distributed in 25% and 72% with group L, respectively, and there were no significant differences in penetration rates between 5.3-mm– and 6.3-mm–thickness groups.