The Journal of Arthroplasty, Volume 30, Issue 6, 1008 - 1013

Performance of Porous Tantalum vs. Titanium Cup in Total Hip Arthroplasty: Randomized Trial with Minimum 10-Year Follow-Up

Wegrzyn, Julien et al.
Hip

Porous tantalum monoblock cups have been proposed to improve survivorship of cementless primary THA. However, there are few direct comparative trials to established implants such as porous-coated titanium cups. 113 patients were randomized into two groups according to the cup: a porous tantalum monoblock cup (TM) or a porous-coated titanium monoblock cup (control). At a mean of 12 years after THA, no implants migrated in both groups. Two TM patients (4%) and 13 control patients (33%) presented with radiolucency around the cup (P < 0.001). In the control group, 1 cup (2%) was revised for aseptic loosening. At 12 years post-implantation, porous tantalum monoblock cups demonstrated 100% survivorship, and significantly less radiolucency as compared to porous-coated titanium monoblock cups.


Link to article