J Orthop Surg Res 14, 437 (2019).

Patient-specific instrumentation improved three-dimensional accuracy in total knee arthroplasty: a comparative radiographic analysis of 1257 total knee arthroplasties

Pauzenberger, L., Munz, M., Brandl, G. et al.
Knee

Background

The purpose of this study was to compare restoration of mechanical limb alignment and three-dimensional component-positioning between conventional and patient-specific instrumentation in total knee arthroplasty.

Methods

Radiographic data of patients undergoing mobile-bearing total knee arthroplasty (n = 1257), using either conventional (n = 442) or patient-specific instrumentation (n = 812), were analyzed. To evaluate accuracy of axis restoration and 3D-component-positioning between conventional and patient-specific instrumentation, absolute deviations from the targeted neutral mechanical limb alignment and planned implant positions were determined. Measurements were performed on standardized coronal long-leg and sagittal knee radiographs. CT-scans were evaluated for accuracy of axial femoral implant rotation. Outliers were defined as deviations from the targeted neutral mechanical axis of > ± 3° or from the intraoperative component-positioning goals of > ± 2°. Deviations greater than ± 5° from set targets were considered to be severe outliers.

Results

Deviations from a neutral mechanical axis (conventional instrumentation: 2.3°± 1.7° vs. patient-specific instrumentation: 1.7°± 1.2°; p < 0.001) and numbers of outliers (conventional instrumentation: 25.8% vs. patient-specific instrumentation: 10.1%; p < 0.001) were significantly lower in the patient-specific instrumentation group. Significantly lower mean deviations and less outliers were detected regarding 3D-component-positioning in the patient-specific instrumentation compared to the conventional instrumentation group (all p < 0.05).

Conclusions

Patient-specific instrumentation prevented from severe limb malalignment and component-positioning outliers (> ± 5° deviation). Use of patient-specific instrumentation proved to be superior to conventional instrumentation in achieving more accurate limb alignment and 3D-component positioning, particularly regarding femoral component rotation. Furthermore, the use of patient-specific instrumentation successfully prevented severe (> 5° deviation) outliers.


Link to article