Background

While patellar resurfacing can affect patellofemoral kinematics, the effect on tibiofemoral kinematics is unknown. We hypothesized that patellar resurfacing would affect tibiofemoral kinematics during deep knee flexion due to biomechanical alteration of the extensor mechanism.

Methods

We performed cruciate-retaining TKA in fresh-frozen human cadaveric knees (N = 5) and recorded fluoroscopic kinematics during deep knee flexion before and after the patellar resurfacing. To simulate deep knee flexion, cadaver knees were tested on a dynamic, quadriceps-driven, closed-kinetic chain simulator based on the Oxford knee rig design under loads equivalent to stair climbing. To measure knee kinematics, a 2-dimensional to 3-dimensional fluoroscopic registration technique was used. Component rotation, varus-valgus angle, and anteroposterior translation of medial and lateral contact points of the femoral component relative to the tibial component were calculated over the range of flexion.

Results

There were no significant differences in femoral component external rotation (before patellar resurfacing: 6.6 ± 2.3°, after patellar resurfacing: 7.2 ± 1.8°, p = 0.36), and less than 1° difference in femorotibial varus-valgus angle between patellar resurfacing and non-resurfacing (p = 0.01). For both conditions, the medial and lateral femorotibial contact points moved posteriorly from 0° to 30° of flexion, but not beyond 30° of flexion. At 10° of flexion, after patellar resurfacing, the medial contact point was more anteriorly located than before patellar resurfacing.

Conclusion

Despite the potential for alteration of the knee extensor biomechanics, patellar resurfacing had minimal effect on tibiofemoral kinematics. Patellar resurfacing, if performed adequately, is unlikely to affect postoperative knee function.