Acta Orthopaedica Scandinavica, 75:4, 464-470

Navigation in total-knee arthroplasty CT-based implantation compared with the conventional technique

Lars Perlick, Holger Bäthis, Markus Tingart, Carsten Perlick & Joachim Grifka
Knee

Background Exact axial limb alignment in total-knee arthroplasty (TKA) is important for a successful outcome. We evaluated the efficiency of computer-assisted implantation in TKA and compared it with the classical surgeon-controlled technique. Patients and methods We implanted 100 TKA using either the computer-assisted technique (50) or the conventional approach (50). There were no significant differences between the groups regarding the preoperative leg deformity. Accuracy of implantation was determined in postoperative long-leg coronal and lateral radiographs. Results A postoperative leg axis between 3° varus and 3° valgus was achieved in 46 patients in the group with computer-assisted implantation and 36 patients in the control group (p=0.01). A significant difference was also seen for the femoral component alignment in frontal plane. No complications influencing the clinical outcome were observed. Interpretation A CT-based navigation system improves the accuracy of TKA, but higher costs and time-consuming planning will mean that its usage is limited to special cases. Additional tools such as ligament balancing, which are presently only available with the CT-free software module, require to be added to the CT-based system.


Link to article