International Orthopaedics February 2009, Volume 33, Issue 1, pp 83–87

Navigated non-image-based positioning of the acetabulum during total hip replacement

Jenny, JY., Boeri, C., Dosch, JC. et al.
Hip

We tested the hypothesis that the non-image-based navigation system used in our department was able to measure accurately the 3D positioning of the acetabular cup of a total hip replacement (THR) and to increase the accuracy of its implantation during THR. We studied 50 consecutive navigated implantations of a THR and compared the intra-operative measurement of the cup by the navigation system to the post-operative measurement by computed tomography (CT) scan. The mean difference between the navigated and CT scan measurements for cup inclination was 2°. The mean difference between the navigated and CT-scan measurements for cup flexion was 4°. These differences were significant but considered to be clinically irrelevant in most cases. A total of 73% of the cases were within the safe zone defined prior to the study. The non-image-based system used allows a precise orientation of the cup during THR.