International Orthopaedics January 2013, Volume 37, Issue 1, pp 15–19

Modular sleeves with ceramic heads in isolated acetabular cup revision in younger patients—laboratory and experimental analysis of suitability and clinical outcomes

Helwig, P., Konstantinidis, L., Hirschmüller, A. et al.
Hip

Purpose

One possibility in hip arthroplasty revisions is to combine a modular ceramic head with an adapter or sleeve in isolated acetabular cup replacement. This study consisted of an experimental part to analyse the reliability of taper damage predictions, and a clinical part to analyse the outcomes of modular ceramic head implantation in a case series of isolated cup replacements.

Methods

Analysis of scratch size on 11 explanted hip stems were examined macroscopically and by stereomicroscopy to classify damage to the conical taper. The second part consisted of a prospective analysis of isolated cup revisions using a modular ceramic head, performed in two orthopedic centres.

Results

Analysis of scratch size on the taper yielded inter-observer correlations of 0.545–0.909; comparison with stereomicroscopic data recordings yielded a moderate correlation, with values between 0.545 and 0.090. The clinical study included 47 isolated acetabular cup revisions involving modular ceramic heads. Ceramic head failure did not occur during the average clinical and radiological observation period of 26 months. Mean Harris Hip Score (HHS) at follow-up was 70 points.

Conclusion

From these results, it can be concluded that damage to the implanted stem taper cannot be reliably predicted intraoperatively. Nevertheless, the clinical outcomes did show that there were no problems with the ceramic heads over the short observation period. The application of modular ceramic heads in younger patients requiring isolated cup replacement requires proper risk-benefit analysis, but is possible and appears to be a safe procedure.