The Journal of Arthroplasty, Volume 29, Issue 8, 1553 - 1558

Modular Necks in Primary Hip Arthroplasty Without Anatomical Deformity: No Clear Benefit on Restoration of Hip Geometry and Dislocation Rate. An Exploratory Study

Gerhardt, Davey M.J.M. et al.
Hip

Modular necks were introduced in total hip arthroplasty (THA) to improve restoration of hip-geometry and reduce dislocation-rates. This presumed advantage was evaluated retrospectively for patients with arthritis in otherwise anatomically normal hips. Restoration of hip-geometry was assessed on preoperative and postoperative calibrated radiographs in 95 consecutive primary THAs with a modular neck design and compared with 95 match controlled THAs with a similar monoblock stem. No significant differences were seen in restoration of body moment arm, leg length and cupangle. Offset restoration revealed a borderline significant difference (P = 0.48) with higher values for the monoblock stem. In both groups 4 dislocations within one year were encountered. In this study modular necks did not reveal a clear benefit in restoring hip geometry and dislocation rate after straightforward THA.


Link to article