The Journal of Arthroplasty, Volume 30, Issue 5, 762 - 765

Midflexion Laxity After Implantation Was Influenced by the Joint Gap Balance Before Implantation in TKA

Minoda, Yukihide et al.
Knee

The relationship between the joint gap before and after implantation in 259 knees during the total knee arthroplasty was investigated using a tensor device which can attach the polyethylene insert trial. Patients were divided into following 3 groups according to the joint gap balance before implantation (flexion joint gap −  extension joint gap); group 1: >1 mm; group 2: −1 to 1 mm, and group 3: <−1 mm. Joint gap after implantation was loose at 30°, 60°, 90°, and 120° of flexion in group 1 and 2, but loose only at 30° of flexion in group 3 (p < 0.01). This study showed that loose flexion joint gap before implantation increased the risk of joint gap laxity after implantation especially at midflexion ranges.


Link to article