The purpose of this study was to characterize the prevalence of taper damage in modular TKA components. One hundred ninety-eight modular components were revised after 3.9 ± 4.2 years of implantation. Modular components were evaluated for fretting corrosion using a semi-quantitative 4-point scoring system. Design features and patient information were assessed as predictors of fretting corrosion damage. Mild-to-severe fretting corrosion (score ≥2) was observed in 94/101 tapers on the modular femoral components and 90/97 tapers on the modular tibial components. Mixed alloy pairs (p = 0.03), taper design (p < 0.001), and component type (p = 0.02) were associated with taper corrosion. The results from this study supported the hypothesis that there is taper corrosion in TKA. However the clinical implications remain unclear.
CoxaPro
> Clinical Library > Welcome to the joint replacement clinical library > Mechanically Assisted Taper Corrosion in Modular TKA
The Journal of Arthroplasty, Volume 29, Issue 9, 205 - 208
Knee
Mechanically Assisted Taper Corrosion in Modular TKA
Arnholt, Christina M. et al.Knee